7.2.25 Communications Port Options 

The 9060Z transmitter has a serial communications port available at terminals 18 to 22. The default 
protocol is for RS‐232 running at 19,200 baud rate with 8 bits, Even parity and 1 stop bit. 

 

Function 

 Options  

60  Serial Interface  

RS‐232 or RS‐485 (Use RS‐485 for MODBUS)  

61  Serial Baud Rate  

9600, 19200, 38400, 57600 (RS‐232 up to 19200 only)  

62    Serial Parity   

 Even, Odd or No Parity  

63  

 MODBUS 
Address   

 Set the MODBUS address for this device between 1 and 246.  
 0 is to disable the MODBUS   

7.2.26 Alarm Log Clearing 

Every alarm that is instigated internally in the transmitter or as a process level is recorded in the alarm 
log  with  the  activation  time,  accepted  time  and  the  cleared  time.  The  last  4000  events  will  be 
recorded and then the oldest alarms will be dropped off as new ones occur. If you would like to delete 
the log recording select CLEAR in SETUP step 64 by using the OPTION keys and press the ENTER key. 
The message “Alarm Log Cleared” will be displayed.