Model 9060Z Manual 

 

29 

 

 

 

SETUP / RUN key 
Pressing  this  key  will  put  the  transmitter  into  one  of  the  menu‐driven  SETUP,  COMMISSIONING  or 
CALIBRATION modes (see chapters 6, 7 and 8 for details). The function of all the keys will then change 
to the functions that they have in the SETUP mode. 

DISPLAY UP / DOWN keys 
The display keys are used to scroll the lower line up and down through the variety of measurements 
that are available on the lower line. For a complete list of options see chapter 6.3.2, Lower Line Items. 

ALARM UP key 
If there is either a new alarm or an active alarm the ALARM UP key can be pressed to examine the 
alarm status. The alarm light will be flashing if there is a new alarm or steady if there is an existing 
alarm. For more details on the alarm mode and keys see chapter 9, Alarms. The SETUP light will flash 
slowly to show that the transmitter is now in the alarm display mode. 

ALARM DOWN key 
When the transmitter is displaying active alarms (the ALARM UP key has been pressed), the ALARM 
DOWN key and the ALARM UP key allow the operator to examine the date / time of the alarm and the 
date / time that the alarm was acknowledged. 
If the transmitter was in RUN mode when this key is pressed it will go into the alarm log display mode. 
See chapter 9 for more details. 

ALARM ACCEPT key 
The ALARM ACCEPT key is used to accept a new alarm (see chapter 9, Alarms). 

GAS 1 / PURGE 1 and GAS 2 / PURGE 2 keys 
These two keys are used to turn on the gas / purge solenoids. When the transmitter is in the manual 
cal / purge mode (Commissioning functions #21 and 30) the solenoid will be activated for as long as 
the key is pressed. When the transmitter is in the auto cal / purge mode the automatic cal / purge 
cycle is started. The cycle can be stopped by pressing the same key again. 

5.3.2 Keypad in SETUP / COMMISSIONING / CALIBRATION modes 

NOTE:  In  order  to  enter  either  COMMISSIONING  or  CALIBRATION  modes,  the  corresponding  DIP 
switch on the 1732‐1 PCB must be turned on. 

From RUN mode, if the SETUP / RUN key is pressed once, the transmitter will go into the SETUP mode. 

From RUN mode, if the SETUP / RUN key is pressed and held for 2 seconds, the transmitter will go into 
the COMMISSIONING mode. 

From RUN mode, if the SETUP / RUN key is pressed and held for 4 seconds, the transmitter will go into 
the CALIBRATION mode. 

The following key functions are then available in all of the above modes. 

SETUP / RUN key 
Pressing the SETUP / RUN key while in any of the set up modes will return the transmitter to the RUN 
mode.