beautypg.com

Model 9060z manual, Important notices – Teledyne 9060Z - Zirconium oxide flue gas oxygen analyzer User Manual

Page 7

background image

Model 9060Z Manual 

 

 

 

 

 

1. IMPORTANT NOTICES 

This manual is intended as a supplement to the 9060Z Operators Manual. It is intended to be used by 
technical personal that are qualified to install, commission, service and calibrate electronic industrial 
control equipment. 
Please read the safety information below before connecting power to the transmitter. 

CAUTION 1 
The  probe  heater  is  supplied  with  mains  voltage.  This  supply  has  electrical  shock  danger  to 
maintenance  personnel.  Always  isolate  the  transmitter  before  working  with  the  probe.  The  EARTH 
wires (green/yellow) from a heated probe must ALWAYS be connected to earth. 

CAUTION 2 
Combustion or atmosphere control systems can be dangerous. Burners must be mechanically set up 
so  that  in  the  case  of  equipment  failure  the  system  cannot  generate  explosive  atmospheres.  This 
danger is normally avoided with flue gas trim systems by adjustment so that in the case of failure the 
appliance will not generate CO in excess of 400 ppm in the flue. The CO level in the flue should be 

measured with a separate CO instrument normally an infrared or fuel cell type. 

CAUTION 3 
 The oxygen probe is heated to over 700°C (1300°F) and is a source of ignition. Since raw fuel leaks can 
occur during burner shutdown, the transmitter has an interlocking relay that removes power from the 
probe heater when the main fuel shut‐off valve power is off. If this configuration does not suit or if it 
is  possible  for  raw  fuel  to  come  into  contact  with  a  hot  oxygen  probe  then  the  Model  9060Z 
transmitter  with  a  heated  probe  will  not  be  safe  in  your  application.  An  unheated  probe  can  be 
utilized in such applications, however the oxygen readings are valid only above 650°C (1200°F). 

CAUTION 4 

The reducing oxygen signal from the transmitter and the associated alarm relay can be used as an 
explosive  warning  or  trip.  This  measurement  assumes  complete  combustion.  If  incomplete 
combustion is possible then this signal will read less reducing and should not be used as an alarm or 
trip. A true excess combustibles analyzer, normally incorporating a catalyst or thermal conductivity 
bridge, would be more appropriate where incomplete combustion is possible. Also read the probe 
electrical shock caution in the probe heater interlock caution in chapter 4.7 

CAUTION 5 

FIL‐3 filter. If the optional FIL‐3 has been fitted to the 9060H probe in this installation, please read the 
Important Notice regarding probe option FIL‐3 on the previous page. 

CAUTION 6 

CAUTION 7 

The  heater  is  supplied  from  the  mains  power  directly,  and  the  temperature  is  controlled  at  720°C 
(1330°F).  The  outside  of  the  process  end  of  the  probe  can get  to  temperatures  that  are  dangerous  to 
touch. Wear insulating gloves when handling a probe that has been on.