beautypg.com

Model 9060z manual – Teledyne 9060Z - Zirconium oxide flue gas oxygen analyzer User Manual

Page 38

background image

Model 9060Z Manual 

 

38 

 

 

 

7.2.8 Flue Pressure Units and Value 

The 9060Z transmitter is capable of fixed or variable input pressure compensation in the calculation of 
oxygen from ‐1 Atm to +3 Atm. If the probe is running in a pressurized environment this value should 
be set to allow  for accurate oxygen measurement. If the flue pressure is constant select 'Fixed" in this 
function and set the pressure units and value in the proceeding menus. If the pressure varies, select 
"Variable", and connect a pressure transducer to terminals 24 & 25. (See section 4.23). Set the range 
of the transducer using a zero and span value in the commissioning menus. 

7.2.9 Temperature Units 

The display on the transmitter can be changed to show temperature in either Celsius or Fahrenheit 
scales. 

7.2.10 Calibration Freezes Outputs 

During gas calibration checks it may be required that the analogue outputs remain frozen at their last 
reading. When cal freezes is enabled both analogue outputs remain frozen until the entire calibration 
cycle has completed. 

7.2.11 Solenoid 1 & 2 Operation 

The  two  solenoid  outputs  can  be  individually  configured  to  perform  gas  calibration  check  or  probe 
purge. 
Some  of  the  menus  between  functions  22  to  28  (31  to  37)  will  not  appear  depending  on  what  has 
been selected in function #20 and #21 (29 and 30). 

7.2.12 Solenoid 1 & 2 Automatic / Manual 

When  set  to  automatic  the  solenoid  will  perform  a  pre‐programmed  gas  calibration  or  purge  cycle. 
When set to manual the solenoid is controlled directly from the corresponding key on the front of the 
case. When set to automatic the calibration check or the purge cycle can be started by pressing the 
corresponding key on the front of the case. It can be stopped by pressing the same key again. 

7.2.13 Solenoid 1 & 2 Start Time 

For automatic gas calibration checking or purge events a start time is specified to allow precise control 
over the timing of such events. Starting at this specified time, the next gas calibration or purge event 
will occur in successive intervals from this time. 

7.2.14 Solenoid 1 & 2 Period 

This option specifies the period between automatic gas calibration or purge events. For periods less 
than 24 hours the period is divisible into 24 hours forcing scheduling of events to occur at the same 
time each day between 1 minute and 7 days. 

The  ‘No  Timed  Operation’  option  accessible  by  scrolling  the  option  below  1  minute  allows  for  pre‐
configured  solenoid  cycles  to  be  triggered  manually  via  the  keypad,  but  without  the  automatic 
scheduling. 

7.2.15 Solenoid 1 & 2 Duration 

Duration of time that an automatic gas calibration or purge event energizes the solenoid