Model 9060Z Manual 

 

28 

 

 

 

 

5.2 OXYGEN DISPLAY UNITS 

The  measured  oxygen  content  displayed  on  the  top  line  of  the  transmitter  for  probe  1  and  on  the 
lower line for probe 2 can be shown in either % or parts per million (ppm). See chapter 6.3.3 Oxygen 
Display Units for how to change the oxygen display. 
The format of the oxygen display changes to maintain the best resolution for the measurement. 
If ‘Oxygen %’ is selected the transmitter will display oxygen in the following format: 

Range   

 

 

Display format 

30.0 to 100.0%  

 

##.#% 

1.00 to 29.99%  

 

##.##% 

0.100 to 0.999%  

 

#.###% 

< 0.100%  

 

 

scientific notation (#.## x 10 ‐ ## % ) 

If  ‘Oxygen  ppm’  is  selected  the  transmitter  will  show  the  oxygen  in  ppm  format  between  10  and 
9,999ppm. Outside of this range the oxygen will be shown in % format described above. 
Display format for ‘Oxygen ppm’: 

Range   

 

 

Display format 

> 10,000ppm (1%)  

 

% format (see ‘Oxygen %’ section above) 

1,000 to 9,999ppm    

#,### ppm 

100.0 to 999.9ppm    

###.# ppm 

10.00 to 99.99ppm    

##.## ppm 

< 10ppm  

 

 

scientific notation (#.## x 10 ‐ ## % ) 

5.3 KEYPAD 

There are eight keys built into the decal on the outside of the door of the 9060Z transmitter. The key 
functions are written in BLACK and WHITE to identify the function of the key in either RUN mode or 
SETUP mode. 
 

Key text  

      RUN mode (WHITE text)  

SETUP mode (BLACK text) 

SETUP / RUN  

Enter SETUP mode  

Return to RUN mode 

DISPLAY / FUNCTION UP  

Display Next Lower Line Item  

Next Function 

DISPLAY / FUNCTION 
DOWN         

Display Prev. Lower Line Item  

Previous Function 

ALARM / OPTION UP  

Next Alarm  

Next / Increment Option 

ALARM / OPTION DOWN  

Previous Alarm  

Previous / Decrement Option 

ALARM ACCEPT / ENTER 

Acknowledge Displayed Alarm  

Save Current Option 

GAS 1 PURGE 1 / SENS IMP 

Gas 1 / Purge 1 manual activate  Check Probe impedance 

GAS 2 PURGE 2 / AUTO CAL   

Gas 2 / Purge 2 manual activate  

Manually Activate Output Calibration 

 

5.3.1 Keypad in RUN mode 

When the transmitter is turned on and has completed its start‐up procedure, it will enter RUN mode 
and enable the keypad. The key functions in this mode are as follows;