Model 9060z manual – Teledyne 9060Z - Zirconium oxide flue gas oxygen analyzer User Manual
Page 28
Model 9060Z Manual
28
5.2 OXYGEN DISPLAY UNITS
The measured oxygen content displayed on the top line of the transmitter for probe 1 and on the
lower line for probe 2 can be shown in either % or parts per million (ppm). See chapter 6.3.3 Oxygen
Display Units for how to change the oxygen display.
The format of the oxygen display changes to maintain the best resolution for the measurement.
If ‘Oxygen %’ is selected the transmitter will display oxygen in the following format:
Range
Display format
30.0 to 100.0%
##.#%
1.00 to 29.99%
##.##%
0.100 to 0.999%
#.###%
< 0.100%
scientific notation (#.## x 10 ‐ ## % )
If ‘Oxygen ppm’ is selected the transmitter will show the oxygen in ppm format between 10 and
9,999ppm. Outside of this range the oxygen will be shown in % format described above.
Display format for ‘Oxygen ppm’:
Range
Display format
> 10,000ppm (1%)
% format (see ‘Oxygen %’ section above)
1,000 to 9,999ppm
#,### ppm
100.0 to 999.9ppm
###.# ppm
10.00 to 99.99ppm
##.## ppm
< 10ppm
scientific notation (#.## x 10 ‐ ## % )
5.3 KEYPAD
There are eight keys built into the decal on the outside of the door of the 9060Z transmitter. The key
functions are written in BLACK and WHITE to identify the function of the key in either RUN mode or
SETUP mode.
Key text
RUN mode (WHITE text)
SETUP mode (BLACK text)
SETUP / RUN
Enter SETUP mode
Return to RUN mode
DISPLAY / FUNCTION UP
Display Next Lower Line Item
Next Function
DISPLAY / FUNCTION
DOWN
Display Prev. Lower Line Item
Previous Function
ALARM / OPTION UP
Next Alarm
Next / Increment Option
ALARM / OPTION DOWN
Previous Alarm
Previous / Decrement Option
ALARM ACCEPT / ENTER
Acknowledge Displayed Alarm
Save Current Option
GAS 1 PURGE 1 / SENS IMP
Gas 1 / Purge 1 manual activate Check Probe impedance
GAS 2 PURGE 2 / AUTO CAL
Gas 2 / Purge 2 manual activate
Manually Activate Output Calibration
5.3.1 Keypad in RUN mode
When the transmitter is turned on and has completed its start‐up procedure, it will enter RUN mode
and enable the keypad. The key functions in this mode are as follows;