Model 9060z manual, 1 hardware specifications – Teledyne 9060Z - Zirconium oxide flue gas oxygen analyzer User Manual
Page 10
Model 9060Z Manual
10
Caution, risk of danger. See additional information in the manual.
3.1 HARDWARE SPECIFICATIONS
3.1.1 TRANSMITTER SPECIFICATIONS
Number of oxygen probes: 2 maximum
Oxygen range:
0‐1% to 0‐100%
Oxygen accuracy:
±1% of actual measured oxygen value with a repeatability of ±0.5% of
the measured value
Thermocouple types:
Type K, J, R and S
Temperature accuracy:
±2°C
Analogue outputs:
0‐20mA or 4‐20mA field selectable
Active outputs (Do NOT loop power these outputs)
Output load:
1000 ohm max
Alarm relays:
4
Alarm relay contacts:
2A/240VAC, 2A/30VDC
(WARNING: Do not use both mains voltage and low voltage connections to adjacent alarm contacts)
Mains voltage supply:
100 to 240VAC, ‐6 +10%, 50/60 Hz
Overvoltage:
Category II (IEC60364‐4‐443)
Power:
5W for controller plus probe power
530W max., 25% duty cycle each probe on 240VAC
110W max., 100% duty cycle each probe on 110VAC
2.5A max
Fuses:
3A, fast blow, 250v, 20x5mm (heater fuses, 2 of)
1A, slow blow, 250v, microfuse (PCB mtg fuse, 1 of)
Environmental Rating:
Operating Temperature ‐25°C to 55°C
Relative Humidity 5% to 95% (non‐condensing)
Altitude
2000m maximum
Degree of Protection:
IP65
IP54 with internal reference air pump
Case Size:
260mm (10.2”) wide, 160mm (6.3”) high, 90mm (3.5”) deep
Weight:
3 Kg (6.6 lbs.)
WARNING: All signal level connections onto the transmitter must be treated as safety extra‐low
voltage (SELV) as defined in the standard IEC61140. Double insulation must be used when
connecting these terminals to systems that might carry high voltage.
3.1.2 STANDARD ‘U’ LENGTHS
9060H
9060UL/UH
250 mm (10”)
500 mm (20”)
350 mm (14”)
750 mm (30”)
500 mm (20”)
1000 mm (40”)
750 mm (30”)
1500 mm (60”)
1000 mm (40”)
1500 mm (60”)
2000 mm (80”)
Ordering Information