Model 9060Z Manual 

 

45 

 

 

 

9. ALARMS 

The 9060Z has 4 alarm relays and a built in alarm annunciator. If an alarm occurs, the ALARM light will 
flash. To find out what the alarm is, press the ALARM up key. 
When the ALARM up key has been pressed, the transmitter goes into the alarm display mode. In this 
mode some of the keys take on a special function. 

Key text  

RUN mode  

ALARM mode  

SETUP / RUN  

*  

Return to RUN mode  

DISPLAY / FUNCTION up  

*  

Next alarm  

DISPLAY / FUNCTION down  

*  

Last alarm  

ALARM / OPTION up  

Enter ALARM display mode  

Alarm activated time  

ALARM / OPTION down  

Enter ALARM log mode  

Alarm acknowledged time 

ALARM ACCEPT / ENTER  

*  

Acknowledge alarm  

GAS 1 PURGE 1 / SENS IMP  

*  

*  

GAS 2 PURGE 2 / AUTO CAL   *  

*  

* This key is not used in the ALARM mode  

The  common  alarm  relay  is  used  to  monitor  faults  within  the  transmitter  and  the  probe.  The  other 
three alarm relays relate to the process gas. All relays have user defined actions. 
If one of the alarm events is removed from the common alarm using COMMISSIONING menu function 
#49, then this alarm event can be programmed for one of the other 3 alarm relays. 
When the alarm mode has been entered, the SETUP light flashes once a second until the SETUP key is 
pressed to return the transmitter to RUN mode. 

All relays have fail‐safe alarm contacts. That is – 

When the transmitter is turned off the contacts are open circuit 
When the transmitter is on but there are no alarms the contacts will be closed 
When there is a current unaccepted alarm event the contacts will be open circuit 
When  there  is  a  current  accepted  alarm  event  the  state  of  the  contacts  will  depend  on  the 
selection in the commissioning menu #50. See chapter 7.2.22, Operation of the alarm relay when 
an alarm is accepted. 

All alarms drive the alarm light on the front door. 

The light will be off if there are no alarms current 
The light will flash if there is a current alarm that has not been acknowledged 
The light will be on steady if there are current alarm(s) that have not been cleared 
The light will flash faster as more alarms occur 

9.1 COMMON ALARMS 

The events that drive the common alarm are – 
1. ‘Probe 1 Heater Fail’ 
2. ‘Probe 2 Heater Fail’ 
A heated probe (model 9060H) has been turned on for more than 20 minutes but is still not up to 
650°C (1200°F) or the probe temperature falls below 650°C. 
3. ‘Probe 1 High Impedance’