Model 9060Z Manual 

 

24 

 

 

 

network  line  from  the  transmitters  is  taken  from  the  middle  of  the  transmitter  network  string,  a 
terminating resistor should be enabled with LK3 at each end of the network line. 

 

The MODBUS™ protocol of the network is –

Baud Rate  

19,200 

Parity  

Even  

Stop Bits  

RS485  

2‐wire (half duplex)  

Mode  

RTU (binary mode)  

For more details see Appendix 5.  
 

 

 

   

 

 

 

 

 

 

 

 

 
4.14 ONNECTING POWE

Before commissioning the probe, sensor or transmitter, read the CAUTION paragraphs at the front of 
this manual. The green & yellow EARTH wire MUST be connected to earth. 

4.15 COMMISSIONING ‐ RUN MODE 

When  the  transmitter  is  turned  on  it  will  go  to  RUN  mode.  The  SET‐UP/RUN  button  will  toggle 
between  the  two  modes.  The  upper  line  of  the  display  will  now  read  ‘%  OXYGEN’.  If  the  probe  or 
sensor temperature is not above 650°C (1200°F), a “Probe Low Temperature” message is flashed on 
the lower line. The probe or sensor temperature can be checked on the lower line of the display. 

4.16 PROBE OR SENSOR CALIBRATION 

The  zirconia  sensor  provides  an  absolute  measurement  of  oxygen  partial  pressure.  There  are  no 
calibration adjustments, apart from ‘Probe 1 Offset’, for the probe. The zirconia sensor EMF is either 
correct or it needs to be replaced. To check that the probe is functioning correctly, first check that the