The  on‐line  auto  purge  and  calibration  check  system  is  optional.  For  details  on  its  operation  see 
chapters 4.24, 4.25 and chapter 11. 
To automatically sense a blocked probe filter, a flow switch should be installed in the ‘purge’ line to 
the probe ‘CAL’ port. It should be adjusted so that it energizes just below the purge flow with new or 
clean filters installed. The flow switch contacts should be connected to terminals 16 & 17 (FLOW SW). 
If the filter is still blocked or partly blocked after an auto purge cycle, the flow switch will not energize 
and will therefore cause a ‘Probe1 (2) Filter Blocked’ alarm. 
After  installation  the  purge/cal  system  should  be  tested  thoroughly  for  leaks.  Any  leaks  can  cause 
significant errors if the flue is at negative pressure. If the flue is at positive pressure, an outward leak 
can cause corrosion in the purge/cal system piping and fittings. 
If probe/filter purging is required but a “Probe1 (2) Filter Blocked” alarm is not required, do not install  
the flow switch but link terminals 16 &17. 

 

4.12 CONNECTING REFERENCE AIR 

For oxygen probes, a 1/4” tube connector on the transmitter should be connected via a nylon, copper 
or stainless steel tube to the ‘REF’ connector on the probe. If two probes are being used, a “T” fitting 
must be supplied to provide reference air supply to both probes. 

4.13 CONNECTING THE TRANSMITTER TO A MODBUS™ NETWORK 

The transmitter can be networked to other transmitters and to a network master. The network uses 
the  transmitter’s  RS485  port.  Up  to  31  transmitters  can  be  connected  to  the  network,  and  can  be 
interrogated by the Network Master. 
NOTE:  Hardware  Protocol  Selection  The  communications  port  of  the  9060Z  can  be  configured  to 
communicate on either RS232 or RS485. If the transmitter is to be used on a MODBUS network, the 
9060Z transmitter must be set to RS485. For further details see chapter 7.2.25, Communication Port 
Options. 
NOTE: Terminating Resistor There is a terminating 100 ohm resistor fitted to the 1732‐1 PCB. Link LK3 
(TERM), in the bottom left‐hand corner of the PCB, is used to connect the terminating resistor. Link 
LK3 must be removed on all transmitters except the transmitter on the end of the network line. If the 

 

Automatic Purge &
Calibration  check 
System 

Wiring 

Schematic 

 

Automatic Purge &
Calibration  check 
System 

Piping 

Schematic