11. GAS CALIBRATION CHECK AND PURGE 

The  oxygen  sensor  that  is  used  in  the  Teledyne  Analytical  Instruments  oxygen  probe  is  extremely 
predictable, stable and reliable. For this reason, the calibration of a Teledyne Analytical Instruments 
oxygen  system  does  not  require  the  use  of  calibration  gases.  However,  all  Teledyne  Analytical 
Instruments oxygen probes have a built in gas connection that does allow the accuracy of the probe to 
be checked in‐situ. This chapter describes the operation of this gas checking system. 
The  9060Z  has  a  timer  and  solenoid  driving  system  that  can  be  configured  to  admit  a  certified 
calibration gas into the probe or an air supply to purge the probe filters through the gas connection. 
Both the calibration gas and the filter purge gas must be piped to the port on the probe labeled “CAL”. 
There are two solenoids drivers in the 9060Z transmitter. They can be used for any combinations of 
gas  checking  and  probe  filter  purging.  The  only  limitation  is  that  if  the  transmitter  is  in  dual  probe 
mode, then solenoid #1 must be wired and plumbed to the probe #1 and solenoid #2 must be wired 
and plumbed to probe #2. 
The transmitter can also be configured to be in a MANUAL or AUTOMATIC purge and gas check mode. 
Function #  

Function Description   

 

Options 

20 (29)  

Solenoid 1 (2) Operation  

 

Calibration Gas / Purge  

√  

√ 

 √ 

21 (30)  

Solenoid 1 (2) Auto / Manual  

 Automatic / Manual    

√  

√  

√ 

22 (31)  

Solenoid 1 (2) Start Time  

 

00:00 to 23:45  

 

 

√  

√ 

23 (32)  

Solenoid 1 (2) Period   

 

1 hour – 7 days  

 

 

√  

√ 

24 (33)  

 Solenoid 1 (2) Duration  

 

1 – 90 seconds  

 

 

√ 

 √ 

25 (34)  

 Solenoid 1 (2) Post Freeze    

5 – 300 seconds  

 

 

√  

√ 

26 (35)  

Oxygen Content Cal Gas 1 (2)  

 0.1 – 20.9%   

 

 

 

 √ 

27 (36)  

Max Cal Gas 1 (2) Positive Error  

0.1% to 3.0%    

 

 

 

√ 

28 (37)  

Max Cal Gas 1 (2) Negative Error  

0.1% to 3.0%    

 

 

 

√ 

 
 

 

11.1 PURGE 

For processes that have particulate present in the gas stream, probe filters are necessary and should 
be back‐purged with sufficient frequency to avoid blocked filters. Select ‘Purge’ in function #20 (29) 
and ‘Automatic’ in function #30 and then set the times in function #22 (31) to #25 (34) The outputs 
can be frozen during purging if selected in COMMISSIONING menu function #19. A flow switch can be 
wired to terminals 16 and 17. If the flow is not sufficient at the end of the purge cycle to close the 
contacts  of  the  flow  switch  a  ‘Probe  1  (2)  Filter  Blocked’  alarm  is  generated.  If  a  flow  switch  is  not 
being used a short circuit should be placed across terminals 16 and 17 to suppress the ‘Probe 1 (2) 
Filter Blocked’ alarm. 

11.2 CALIBRATION GAS 

If  it  is  required  to  automatically  check  the  oxygen  probe  and  transmitter  on  line,  select  ‘Calibration 
Gas’ in function #20 (29) and ‘Automatic’ in function #30 and then set the times in function #22 (31) 
to #25 (34) and the gas levels in function #26 (35) to 28 (37). If the transmitter measurement does not 
fall  within  the  limits  set  in  function  #26  (35)  to  28  (37)  a  ‘Gas1  (2)  Calibration  Error’  alarm  will  be 
generated. If the transmitter has ‘Automatic’ selected in function #21 (30) and one of the GAS/PURGE 
keys is pressed in RUN mode, the automatic cycle will be started and the alarms will be checked. The 
cycle can be stopped at any time by pressing the same key during the cycle. 

Functions available if either Calibration Gas or Purge and Manual are selected

 

Functions available if Purge and Automatic are selected 

Functions available if Calibration Gas and Automatic are selected