Model 9060z manual – Teledyne 9060Z - Zirconium oxide flue gas oxygen analyzer User Manual
Page 48
Model 9060Z Manual
48
level. With heated probes this relay will be de‐energised while the probe is heating up from ambient,
making the contacts open circuit.
30. ‘Cal 1 in Progress’
31. ‘Cal 2 in Progress’
A calibration check is occurring, either manual or automatic mode.
32. ‘Purge 1 in Progress’
33. ‘Purge 2 in Progress’
A probe purge is occurring, either manual or automatic mode.
9.3 ALARM RELAY OPTIONS
The three process alarm relays (relays 1 to 3) are user defined. The relay action will depend on the
selections made in the COMMISSIONING menu functions #45 to #48.
The relay contacts are all designed to be “Fail Safe”. That is, they are
open when the transmitter power is off
closed when there is NOT an alarm condition
open when an alarm occurs
The contact will close circuit again (relay energised) when the alarm condition is acknowledged.
Choose to have the relays react to any or all of the following alarms or warnings –
Oxygen on probe 1 Low
Alarm
Oxygen on probe 2 Low
Alarm
Oxygen on probe 1 Very Low
Alarm
Oxygen on probe 2 Very Low
Alarm
Oxygen on probe 1 High
Alarm
Oxygen on probe 2 High
Alarm
Oxygen Deviation High
Alarm
Probe 1 Temperature Low
Warning
Probe 2 Temperature Low
Warning
Calibration check on probe 1 in
Progress
Warning
Calibration check on probe 2 in
Progress
Warning
Purge on probe 1 in Progress
Warning
Purge on probe 2 in Progress
Warning
In addition to the above process alarms that can activate the process alarm relays, any of the common
alarm relay events that have been taken off the common alarm list in COMMISSIONING menu
function #49 will appear on the process alarm relay lists in function #46 to #48.