beautypg.com

Model 9060z manual – Teledyne 9060Z - Zirconium oxide flue gas oxygen analyzer User Manual

Page 58

background image

Model 9060Z Manual 

 

58 

 

 

 

APPENDIX 2, PROBE EMF TABLES 

ZIRCONIA OXYGEN SENSOR OUTPUT (mV) PROBE TYPE 9060H, SENSOR  

OXYGEN 

%   

Probe EMF  

@ 720°C (1320°F)  

OXYGEN 

%  

Probe EMF 

@ 720°C (1320°F) 

20.95 

0

10

15.93

20.5 

0.517

9.5

17.033

20 

1.046

9

18.196

19.5 

1.589

8.5

19.426

19 

2.147

8

20.73

18.5 

2.719

7.5

22.12

18 

3.306

7

23.607

17.5 

3.911

6.5

25.204

17 

4.533

6

26.93

16.5 

5.173

5.5

28.808

16 

5.834

5

30.867

15.5 

6.515

4.5

33.145

15 

7.219

4

35.695

14.5 

7.947

3.5

38.59

14 

8.7

3

41.94

13.5 

9.481

2.5

45.913

13 

10.292

2

50.797

12.5 

11.134

1.5

57.135

12 

12.011

0.9

66.182

11.5 

12.925

0.4

82.168

11 

13.881

0.2

99.518

10.5 

14.881

0.1

114.347

'K' Type TC 29.965 mV @ 720°C (1320°F) 

These tables are based on the Nernst equation: O2 
 

EMF 

=

 

.

.

 

Where T is temperature° K