beautypg.com

Model 9060z manual – Teledyne 9060Z - Zirconium oxide flue gas oxygen analyzer User Manual

Page 47

background image

Model 9060Z Manual 

 

47 

 

 

 

19. ‘Gas 1 Calibration Error’ 
20. ‘Gas 2 Calibration Error’ 
At the end of an automatic gas calibration check cycle the oxygen reading was outside the values set 
in the COMMISSIONING menu #26 to 28 (#35 to 37). This alarm will not reset until the next calibration 
check cycle. The cycle can be initiated manually or automatically. 

21. ‘BBRAM Fail’ 
All of the setup options are held in the battery backed memory (BBRAM). This is the battery shaped 
device at the bottom centre of the 1732‐1 PCB labeled MEM1. This alarm will occur when this device 
fails and will need to be replaced. 

9.2 SELECTABLE PROCESS ALARMS 

The alarm relays 1, 2 and 3 are generally used to transmit a gas related alarm event. Any or all of the 
following functions can be selected for each relay. 
The trip levels and the delay times are set is in the COMMISSIONING menu. 

NOTE: The process alarms will only be activated if they are enabled in the SETUP function 6. See 
chapter 6.3.5. 

21. ‘Oxygen 1 Very Low’ 
22. ‘Oxygen 2 Very Low’ 
The  measured  oxygen  level  on  the  indicated  probe  has  been  below  the  trip  level  shown  in 
COMMISSIONING menu function #38 for longer than the delay time shown in function #39. 

23. ‘Oxygen 1 Low’ 
24. ‘Oxygen 2 Low’ 
The  measured  oxygen  level  on  the  indicated  probe  has  been  below  the  trip  level  shown  in 
COMMISSIONING menu function #40 for longer than the delay time shown in function #41. 

25. ‘Oxygen 1 High’ 
26. ‘Oxygen 2 High’ 
The  measured  oxygen  level  on  the  indicated  probe  has  been  above  the  trip  level  shown  in 
COMMISSIONING menu function #42 for longer than the delay time shown in function #43. 

27. ‘Oxygen Deviation’ 
The difference between the oxygen level measured on probe #1 and the oxygen level measured on 
probe #2 is greater than the trip level shown in COMMISSIONING menu function #44 for longer than 
the delay time shown in function #45. 

28. ‘Probe 1 Temperature Low’ 
29. ‘Probe 2 Temperature Low’ 
The probe temperature is under 650°C (1200°F). The oxygen reading is therefore invalid. If the probe 
heater  has  been  on  for  more  than  20  minutes  and  the  temperature  is  less  than  650°C  (1200°F)  a 
‘Probe # Heater Fail’ alarm will occur. 

NOTE: The ‘Probe # Temperature Low’ relay function is used with unheated probes to indicate oxygen 
reading is invalid (the probe is below 650°C (1200°F) ), in case the process temperature falls below this