Model 9060z manual – Teledyne 9060Z - Zirconium oxide flue gas oxygen analyzer User Manual
Page 11
Model 9060Z Manual
11
Probe insertion length (from process end of mounting thread to probe sensing tip.
3.1.3 SERIES 9060H PROBES SPECIFICATIONS
MODEL
9060H
9060UL/UH
Application
Combustion flue
Combustion flue gases
gases below
above 700°C (1290°F) with
900°C (1650°F)
no contaminants.
Refer to note 1
eg. natural gas, light oils
Temperature Range
0 to 900°C. Refer to note 2
(32 to 1650°F)
700 to 1400°C
(1470 to 2550°F)
Length
250 to 2000 mm
500 to 1500 mm
(10” to 80”)
(20” to 60”)
Process Connection
1 ½ BSP " or NPT
¾ BSP " or NPT
Electrical Connection
Weatherproof plug‐in connector or optional screw terminals. The plug
connector can be supplied with the cable.
Cable
Order a specific length with the transmitter
Heater
Yes
No
Thermocouple
K, integral
R, integral
Response Time
Typically < 4 secs.
Typically < 1 sec
Head Temperature
‐25 to 100°C (‐15 to 210°F) with weatherproof connector
‐25 to 150°C (‐15 to 300°F) with screw terminals
Reference Gas
Ambient air, 50 to 150 cc/min (3 to 9 scim). Pump can be supplied with
transmitter.
Ref Air Connection
1/4 NPT "
Integral air line in probe cable.
Barbed fitting to 3/16" ID PVC tube.
Filter
Removable sintered titanium alloy particulate filter, 30 micron standard,
optional 15 micron available. Refer to note 2
Calibration Check Gas
Connection
1/8 NPT female "
1/8 NPT female "
Weight
2 kg (4.4 lb) plus
165 g (5.8 oz) / 100 mm
(4”) length
1 kg (2.2 lb) plus
100 g (3.5 oz) / 100 mm
(4”) length
Notes: 1. Care must be taken to avoid contact with explosive or inflammable gases with 9060H
heated probes when hot. Teledyne Analytical Instruments transmitters have built in safety
protection.
2. Process gas temperature must be below 800°C if the filters are fitted.
Please contact factory for corrosives other than sulfur or zinc. We can provide test materials to try
in your atmosphere.