13. TROUBLESHOOTING 

The Teledyne Analytical Instruments oxygen transmitter has proved an extremely reliable instrument. 
The trouble shooting guide here is based on an analysis of the potential problem that may occur after 
many  years  of  operation  in  the  field.  A  current  list  of  problems  and  solutions  can  be  found  on  the 
Teledyne Analytical Instruments web site at 

www.teledyne‐ai.com

 when any are identified. 

13.1 FIRST APPROACH 

The probe will not heat up. 
Turn  the  power  off,  remove  and  check  2  x  20mm  glass  fuses.  If  either  of  them  is  blown,  check  the 
probe heater resistance between the two white wires that come from the probe. It should be 110 +/‐ 
15 ohms, and both wires should be open circuit to earth. If not, replace the probe. 
Is the ‘B’ shown on the bottom left hand corner of the display? If not, check that the burner is enabled 
with  either  a  link  between  terminals  10&11,  BURNER  INPUT,  or  is  enabled  with  a  connection  to  a 
voltage‐free contact from the main fuel valve to these terminals. 

The display is blank and there is no backlight on. 
Measure the power supply voltages at the test points labeled COMM and +5v at the right hand side of 
the main PCB 1732‐1. It should be 5 +/‐ 0.15v. 
Turn  off  the  power  and  remove  the  main  shield.  Check  that  the  plug‐in  fuse  FS1  has  not  blown.  It 
should only be replaced with a 1A, 250v fuse if it has blown. 
Is the mains power being supplied to terminals 36&37 and is it between 100 and 240VAC? If may be 
necessary to replace the switch mode power supply, PS5, APC‐5S. 

An ‘ADC Calibration Fail’ alarm has occurred. 
Turn the transmitter off and remove the main shield. Turn the transmitter back on and measure the 
voltages  on  the  ‘Acom’  test  points.  With  the  negative  lead  on  the  centre  terminal  the  other  2 
terminals should measure 12 +/‐0.3v. If not, replace the small DC‐DC converter PS4. 

An ‘Output 1 (2) Failure’ alarm has occurred. 
Turn the transmitter off and remove the main shield. Turn the transmitter back on and measure the 
voltages  on  the  ‘D1com’  test  points.  With  the  negative  lead  on  the  centre  terminal  the  other  2 
terminals should measure 12 +/‐0.3v. If not, replace the small DC‐DC converter PS4. 

A ‘Heater 1 (2) Failure’ alarm has occurred. 
The software has found that the SSR4 has failed. If only one probe is being used and you need to gets 
the transmitter back working quickly, use the CALIBRATION menu function 18, Heater SSR Select, to 
change the heater output terminals. See chapter 8.2.8, Heater SSR fault correction. 

A ‘BBRAM Fail’ alarm has occurred 
Replace the BBRAM, MEM1 on the 1732‐1 main PCB. 

13.2 DETAILED FAULT ANALYSIS 

The  9060Z  transmitter  has  a  diagnostics  mode  built  into  the  software.  This  mode  allows  detailed 
analysis of the hardware of the transmitter, but does require a level of competence in electronics. 
The diagnostics mode is selectable by turning the ‘Diagnostics’ DIP switch to ON and then turn on the 
power.  There  is  a  separate  Diagnostics  Mode  Manual  available  that  describes  its  use.  Ask  Teledyne 
Analytical Instruments for more details.