beautypg.com

Model 9060z manual – Teledyne 9060Z - Zirconium oxide flue gas oxygen analyzer User Manual

Page 25

background image

Model 9060Z Manual 

 

25 

 

 

 

high  impedance  alarm,  ‘Probe  1  (2)  High  Impedance’,  is  not  active.  The  actual  impedance  can  be 
displayed on the lower line. It should be less than 9 KΩ at 720°C (1320°F). 
Once it has been established that the impedance is normal, the offset may be set using the millivolt 
value marked on the oxygen probe. See chapter 6.3.1, Probe Offset. The probe offset can be tested on 
site. A small flow of air must be admitted to both the ‘REF’ and ‘CAL’ ports when testing the probe 
offset. If the probe is in the process, the air must fully purge the probe sensor without interference 
from the process gas sample. Teledyne Analytical Instruments probes can easily achieve this with or 
without a probe filter and a gas flow of only 1 to 5 SLPM (2 to 10 SCFH) for a 9060H probe and up to 
20 SLPM (40 SCFH) for an unheated probe. When a stable oxygen reading has been achieved, read the 
‘Probe EMF’. Enter this value into the set‐up menu #1 for the probe 1 (set‐up #2 for the second probe 
if it is installed). 

4.17 FILTER PURGING 

 

Purging  probe  filters  is  controlled  from  the  ‘GAS/PURGE’  buttons  on  the  transmitter  when  in  ‘RUN’ 
mode.  If  ‘Automatic’  has  been  selected  in  either  ‘Sol.  1  Auto/Man’  or  ‘Sol.  2  Auto/Man’  in 
Commissioning  Menus  21  or  30,  pressing  the  GAS/PURGE  button  will  start  the  automatic  cycle. 
Pressing  the  button  again  will  cancel  the  auto  purge  cycle.  If  ‘Automatic’  was  not  enabled,  the 
solenoid will only stay open for as long as the button is pressed. Gradually adjust the purge air supply 
regulator,  increasing  the  pressure  until  sufficient  flow  is  obtained  to  clear  the  filter.  This  is  best 
checked  with  a  dirty  filter  after  a  period  of  operation,  by  withdrawing  the  probe  from  service  and 
watching  any  build  up  on  the  filter  being  blown  off  at  the  set  pressure.  Normally  30  kPa  (5  psi)  is 
adequate but the air pressure may be set as high as 100 kPa (15 psi). 

4.18 CALIBRATION GAS CHECK 

If the installation has a filter purge facility, set this up first. Refer to the previous paragraph. Press the 
‘GAS/PURGE 1 or ‘GAS/PURGE 2’ button while in ‘SET UP’ mode to obtain a reasonable flow through 
the calibration check gas flow meter. If air is being used as a calibration check gas, use the air from the 
regulator for filter purge. Then, when setting up a gas for calibration checking, set the pressure from 
the calibration gas cylinder so that  it  is the same as the pressure set on the air regulator. Then the 
setting on the rotameter / flow regulator will be the same as that for the airflow. The flow required is 
1 to 5 SLPM (2 to 10 SCFH)  for a 9060H probe and up to 20 SLPM (40 SCFH) for an unheated probe. 
Air  is  not  the  best  gas  for calibration checking on a zirconia sensor.  The output  of  a zirconia sensor 
with air on both sides of the sensor is zero millivolts. It is better to choose a gas value which provides 
a reasonable output from the sensor and which is near to the process oxygen level. A cylinder with 2% 
oxygen  in  nitrogen  is  a  commonly  used  calibration  gas.  The  maximum  pressure  on  the  calibration 
check gas cylinder regulators is 100 kPa (15 psi). 

Note: If ‘Dual probe’ was selected in commissioning menu 4, ‘Cal Gas 2’ must be connected to probe 2. 

4.19 DUST IN THE FLUE GAS 
 

 

 

 

For heated probes the preferred method of mounting for dust‐laden applications is facing vertically 

For  unheated  probes  with  no  filter,  entrained  solids  or  dust  in  the  flue  gas  does  not  present  a 
problem unless the dust, when settled, is not porous. Allow the dust in the process to build up on 
the probe. It will form a porous layer slowing the response time. To avoid mechanical abrasion of 
the  electrode  material  in  installations  with  unheated  oxygen  probes,  pack  ‘SAFFIL’  or  equivalent 
alumina based ceramic fiber in the sensing holes to protect the electrode. Do not use silica based 
ceramic fibers such as ‘KAOWOOL’, which can attack the electrode at high temperatures. Once the 
dust has built up the response time of the probe will be slower.