Model 9060z manual – Teledyne 9060Z - Zirconium oxide flue gas oxygen analyzer User Manual

Page 30

background image

Model 9060Z Manual 

 

30 

 

 

 

FUNCTION up and FUNCTION down keys 
These two keys allow for the selection of the setup function. A function summary table is found at the 
start of the relevant SETUP, COMMISSIONING or CALIBRATION mode chapter. 

OPTION up and OPTION down keys 
These  two  keys  allow  for  changing  the  option  for  the  selected  function.  A  list/range  of  options  for 
each  function  is  found  in  the  function  summary  table  at  the  start  of  the  relevant  SETUP, 
COMMISSIONING or CALIBRATION mode chapter. 

ENTER key 
The ENTER key sets the currently displayed option and stores the value in non‐volatile memory to be 
retrieved  on  device  start  up.  If  the  ENTER  key  is  not  pressed  when  a  new  option  is  chosen,  the 
previous option will be retained. 

Probe impedance key  
When this key is pressed the transmitter will measure the impedance of oxygen sensor(s) attached to 
the transmitter. If the burner is not enabled (terminals 10 & 11) or the probe temperature is below 
700°C (1292°F) impedance checking will not be performed. A ‘Z’ will be seen in the bottom left hand 
corner of the display as the test is performed. 

Auto calibrate key 
When this key is pressed the transmitter will calibrate the analogue output channels that are set to 
auto  calibration  (see  chapter  10).  This  is  performed  by  directing  the  output  current  away  from  the 
output terminals (terminals 12&13 and 14&15) back to an analogue input to the transmitter in order 
to calculate a zero and span calibration factor for each of the output channels. 
During this process normal output to the analogue channels will be interrupted sending the outputs to 
zero mA.