10. INSTRUMENT CALIBRATION 

10.1 CALIBRATION SUMMARY 

The 9060Z oxygen transmitter has a self calibration and diagnostic system built into the hardware and 
software.  Once  the  reference  voltages  have  been  set  into  memory  the  self  calibration  system 
maintains the calibration of the transmitter. An automatic update of the zero and span is done every 
minute. 
The 4‐20mA outputs can be automatically calibrated with the press of the AUTO CAL key or manually 
fine tuned. 
One entry for each probe offset is  required to optimise the calibration of the oxygen measurement 
although  an  automatic  system  calibration  check  can  be  programmed  into  the  transmitter  using 
certified gasses. 

10.1.1 CALIBRATION OF THE INPUTS 

The calibration of the analogue inputs is based on the voltage of a temperature compensated voltage 
reference integrated circuit. There are 4 voltages generated from the standard reference voltage. 
They will vary by about 1% from one transmitter to another but can be trimmed by setting the actual 
voltages into CALIBRATION menu functions #1 to 4. 

The calibration should be done 30 minutes or more after the instrument has been on, approximately 
once every year. The calibration constants are retained in battery backed memory unless a ‘COLD 
START’ is performed. See chapter 10.2, Cold Start. 
Connect a 3 1/2 digit multimeter negative lead to the test point marked ‘COM’ in the centre of the 
1732‐1  PCB  (labeled  ‘V‐REFS’).  Measure  the  four  voltages  on  the  test  point  marked  1  to  4  with  the 

positive lead of the multimeter. Enter the measured values in the CALIBRATION menu functions #1 to 
4. Whenever new values are entered, the D/A Section should be re‐calibrated. See chapter 10.1.2. 

10.1.2 CALIBRATION OF THE OUTPUTS 

The easiest way to calibrate the outputs is to select ‘Auto Calibrated’ in CALIBRATION menu functions 
#5 and 8. If this is selected the outputs will be directed away from the output terminals and back into 
an  analogue  input  of  the  transmitter.  The  outputs  are  then  tested  and  a  zero  and  span  calibration 
factor  is  recorded.  If  a  more  accurate  calibration  is  required  select  ‘Manually  Calibrated’  in 
CALIBRATION menu functions #5 and 8. This will inhibit the automatic calibration system overwriting 
the calibration factors. However, the zero and span factors will need to be manually set. To set the 
calibration factors use the following steps for each output: 

1. Select CALIBRATION menu function #5 (8) and set the option to ‘Calibrate 4mA’. 
2. Use the FUNCTION up key to go to function #6 (9). 
3. Set this to 4.00mA and press the ENTER key. 
4. Measure the actual output current with a digital multimeter. 
5. Use  the  OPTION  keys  to  set  the  option  to  the  value  read  on  the  digital  multimeter  and  press 

ENTER. 

6. The output should go to 4.00mA. The value in function #6 (9) can be fine tuned if necessary. 
7. Go back to function #5 (8) and select ‘Calibrate 20mA’. 
8. Go up to function #7 (10) and set the option value to 20.00mA. 
9. Measure the actual output current with a digital multimeter. 
10. Use  the  OPTION  keys  to  set  the  option  to  the  value  read  on  the  digital  multimeter  and  press 

ENTER.