5. DISPLAY AND KEYPAD
Operator interaction with the oxygen transmitter is done through a graphical display and 8 keys on
the front of the case. The five LED indicators are used to show the status of the transmitter and to
alert the operator of any errors.
Each of the keys has a dual function;
The BLACK text printed on the key is the function while the transmitter is in the RUN mode
The WHITE text printed on the key is the function in one of the three configuring menus, (SETUP,
COMMISSIONING and CALIBRATION)
The menu driven modes are accessed by pressing, and in some cases holding down, the SETUP key
(see chapters 6, 7 and 8 for details of these modes). The transmitter will return to the RUN mode
when the SETUP key is pressed again, or if 60 seconds has elapsed since the last key was pressed
.
5.1 Graphical Display
In RUN mode the 9060Z shows the first oxygen probe measurement in large characters at the top of
the display and a user selectable lower line in smaller characters below. Other items on the display
include the activity heartbeat indicator in the top right corner, a row of single letter action indicators
in the bottom left corner and the current time in the bottom right corner.
In SETUP mode the RUN display is replaced with a menu driven interface. All other functions of the
transmitter (reading, calculating and transmitting etc) aside from the keypad and display will continue
to operate as normal.