Model 9060z manual, Setup mode – Teledyne 9060Z - Zirconium oxide flue gas oxygen analyzer User Manual
Page 31
Model 9060Z Manual
31
6. SETUP MODE
The SETUP mode is mode is accessed by pressing the SETUP key momentarily so the words ‘Setup
Mode’ appear at the bottom of the display. The transmitter will return to the RUN mode when the
SETUP key is pressed again or 60 seconds after the last key press.
6.1 FUNCTION SUMMARY TABLE
When the transmitter is in the SETUP mode the SETUP light will be on and the words “Setup Menu”
will be shown at the bottom of the display.
The following table shows the SETUP functions:
Menu #
Function name
Range
Default value
1
Probe 1 Offset
‐6.0 to +6.0mV
0.0mV
2
Probe 2 Offset
‐6.0 to +6.0mV
0.0mV
3
Lower Line Items
See Chapter 6.3.2
4
Oxygen Display Units
Oxygen % / Oxygen ppm
Oxygen %
5
Damping Factor
No Damping to 10 Samples
5 Samples
6
Process Alarms
Enabled / Disabled
Disabled
6.2 CHANGING OPTIONS
The purpose of having a menu driven SETUP mode is to allow for configuration of the transmitter
using the graphical display and keypad.
Once an option is changed and entered using the ENTER key that value immediately becomes active.
It is also stored into the non‐volatile memory within the BBRAM and will be loaded again at device
start‐up. To reset the transmitter configuration defaults see chapter 10.2, Cold Start.
To change an option in the SETUP menu system:
1. Select the SETUP mode by pressing the SETUP / RUN key once. The SETUP light will come on and
the display will have the format shown below. The operations of the keys are now the operations
written in WHITE on the keypad. The menu name is written at the bottom of the display.
2. When the SETUP mode has been selected the required function can be found by using the
FUNCTION UP and FUNCTION DOWN keys.
The options available for that function can be seen by using the OPTION UP & OPTION DOWN
keys.
3. When the required option is on the display press the ENTER key to save that value.
When finished, press the SETUP / RUN key to return to the RUN mode.