Micromod MOD: 30MLTraining Manual User Manual
Page 154

Training Manual
Debug Lab
• Type R STATE in the command field and then click on the Enter button.
Figure 9 .11.
Instrument Status
• The command you entered will be echoed on the main window above, followed by
the response from the instrument.
3.
Read diagnostics:
• Type R DIAGS in the command field and then click on Enter. Instrument
diagnostics will be displayed as shown in the next figure. You may get more
diagnostic errors or none.
Refer to MOD 30ML / Modcell / ViZapp Designer documentation or the in-program
Help for syntax, commands and readable / writeable attributes listing.
Some common commands are listed below:
ACK_SD
Acknowledge shutdown fault
CLR_Q
Clear event queue
ACK_ALL
Ack all diagnostics, process alarms & notification messages
R VERSION
Read instrument version
R Q
Read instrument event queue
R DIAGS
Read instrument diagnostics
Commands are not case-sensitive. When reading/writing attributes, use the block type
with occurrence number, as shown in the CRF file.