Quick start guide – B&K Precision 2542B-GEN - Manual User Manual
Page 204
Quick Start Guide
204
Measure the phase difference between two signals of the
same frequency under X-Y display mode.
Connect a sine wave signal to CH1 and a sine wave
signal of the same frequency but out of phase to CH2.
Press horizontal MENU key to display the Horizontal
menu.
Press X-Y softkey to select X-Y display mode
Center the signal on the display with the vertical control
knob of each channel.
Use the vertical scale control knob of each channel to
expand the signal for convenient view.
Press the CURSOR key to display the CURSOR menu.
Press Mode softkey to select the Manual mode.
Press Source softkey to select CH2.
Press Type softkey to select the Voltage type.
Press
Y1--/
Y2—softkey or press the entry knob to
select Y1 cursor.
Rotate the entry knob
to move the Y1 cursor to the
top of the signal.
Press
Y1--/
Y2—softkey or press the entry knob to
select Y2 cursor.
Rotate the entry knob
to move the Y2 cursor to the
bottom of the signal.
ΔY displayed in the softkey area is the voltage difference
D (or 2B) between Y1 and Y2.