Xylem photoLab 6100 VIS User Manual
Page 64

Operation
photoLab
®
6100 VIS
64
ba75847e01
08/2009
Variant 1:
Measure standard
solutions
1
Select and confirm Measure
standard solutions.
2
Enter and confirm details of the
standard solutions (optional).
3
Using [Next], accept all entries on
the page and switch to the next
page.
The table for the measurement of
standard solutions pops up.
In the first two lines of the table,
the two value pairs (measuring
points) that are at least required
for a calibration are already
prepared (reagent blank value E0
and any further nominal value).
4
Create further values pairs with
[Add] as necessary.
You can delete a highlighted value
pair with [Delete].
5
In the Target value column, enter
the nominal values of the
individual standard solutions.
Measuring the standard solutions:
6
Using the arrow keys <
▲><▼>
and <
W><X>, navigate to the
relevant input field in the
Absorbance column and press
Edit method
04/16/07 9:52
Standard ID
Standard manufacturer
Back
Next
Edit method
04/16/07 9:52
Target value
Absorbance
E0
0.000
1
Back
Add
Delete
Next
Edit method
04/16/07 9:52
Target value
Absorbance
E0
0.000
1
0.300
2
0.600
3
1.000
Back
Add
Delete
Next
Edit method
04/16/07 9:52
Target value
Absorbance
E0
0.000
1
0.300
2
0.600
3
1.000
Back
Add
Delete
Next