Xylem photoLab 6100 VIS User Manual
Page 63

photoLab
®
6100 VIS
Operation
63
ba75847e01
08/2009
How to program
user-defined
methods
Note
During the following proceeding, you can return to the previous page at any
time with [Back], e. g. if you want to correct entries, add further value pairs or
eliminate outliers.
1
Enter the general method data
here. The next available method
number is already entered as the
number.
You have the following options
when filling out the input fields:
- Fill out all empty input fields one
after the other
- Using [Method list], select an
already existing method as a
model, give it a new method
number and adjust the entries
- Using [Method list], select an
existing method in order to
change it (without changing the
number).
- You can delete the method
completely with [Delete].
2
Select the menu item, Calibration
curve. Select the method for the
determination of the calibration
line. The following variants can be
selected:
- Measure standard solutions
- Enter value pairs
- Enter formula
3
Using [Next], accept all entries on
the page and switch to the next
page.
– [Setup]
– New method
Edit method
04/16/07 9:52
Number
1001
Designation
Nitrite
Version
01
Wavelength
525
Cell
10 mm
Citation form
NO2-N
Unit
mg/l
Resolution
0.001
Calibration curve
Measure standard solutions
Method list
Delete
Next