2 error causes and remedies, Error causes and remedies – Xylem photoLab 6100 VIS User Manual

Page 156

background image

What to do if ...

photoLab

®

6100 VIS

156

ba75847e01

08/2009

6.2

Error causes and remedies

Instrument does not

react to keystroke

Acoustic signal on

keystroke

Measuring range

undercut or

exceeded

Note
In Concentration mode you can display the current absorbance value as
an additional information ([Setup]/Display absorbance, see also section
4.5.6)
.

Self-test does not

start.

The instrument

displays

Please remove cell

Cause

Remedy

– Software error

– Operating condition undefined or

EMC load unallowed

– Processor reset:

Press the and
key simultaneously.

Cause

Remedy

– The key does not have any

function in the current operating
state

– Press a different key

Cause

Remedy

– Method not suitable

– Select method with suitable

measuring range

– Dilute the sample

Cause

Remedy

– Cell is inserted

– Remove the cell

– Press the key.

– Foreign object is in the cell shaft

– Remove foreign object

– Press the key.

– The instrument has to carry out an

automatic balancing process for
the rectangular cell recognition

– Press the key.

– The rectangular cell shaft is con-

taminated

– Clean the cell shaft

– Restart the instrument

– If necessary, confirm the Please

remove cell message with
.

– Instrument defective

– Return instrument to service

department