Carrying out the
AQA2 check for a
method
Test record
A test record is displayed after the check. It can be printed and stored as a
file (in the internal DataB folder or USB memory device at the USB-A
connection, see section 4.11.1).
Sample printout:
Note
Later you can view the last AQA2 test records for all methods monitored with
AQA2 under AQA2 info.
1
Carry out the check like a normal
measurement (see section 4.5.1
to 4.5.3).
2
Insert the cell or
start measurement with
After the measurement is
completed, the result and its
evaluation are displayed.
If the check failed, it is possible to
repeat the measurement.
If the check was successful, the
AQA2 check function is finished.
– AQA2 check
AQA2 check
04/16/07 9:52
Target value
2.00
To start measurement,
insert cell or press
3: A6/25
NH
4
-N
16 mm
0.20 - 8.00 mg/l
SpectroFlex 6100
09130512 1.30-WTW-1.60 Administrator
AQA2
OK
Protocol ID
32
Executed by:
Administrator
Executed
21.05.2007
Valid until:
13.08.2007
Method
6: P6/25 PO4-P
Standard ID
CC10 OC557775
Target value
0.80 +- 0.08 mg/l
Measured value
0.84
mg/l