Sample Experiments

Palm-SPEC Operating Instructions

41

Application Tips

If the signal you collect is saturating the spectrometer (intensity greater than 4000 counts), you can decrease the
light level on scale in scope mode by:

Decreasing the integration time

Attenuating the light going into the spectrometer

Using a smaller diameter fiber (when using an external light source)

Using a neutral density filter with the correct optical density (when using an external light source)

If the signal you collect has too little light, you can increase the light level on scale in scope mode by:

Increasing the integration time

Using a larger diameter fiber (when using an external light source)

Removing any optical filters (if using an external light source)