A: Experiment Tutorials

Procedure

Perform the following steps to take reflection measurements using OOIBase32:

1. Place OOIBase32 in scope mode by clicking the Scope Mode icon on the toolbar, or by selecting

Spectrum | Scope Mode from the menu bar.

2. Ensure that the entire signal is on scale. The intensity of the reference signal should peak at about

3500 counts.

3. Take a reference spectrum with the WS-1 Diffuse Reflectance Standard or the STAN-SSH High-

reflectivity Reference Standard. You must take a reference spectrum before measuring reflection.

Click the Store Reference spectrum icon on the toolbar or select Spectrum | Store Reference
from the menu bar to store the reference. This command merely stores a reference spectrum in
memory. You must select File | Save | Reference from the menu bar to permanently save the
spectrum to disk.

4. Block the light path to the spectrometer. Then, take a dark spectrum by clicking the Store Dark

Spectrum icon on the toolbar or by selecting Spectrum | Store Dark from the menu bar. This
command merely stores a dark spectrum in memory. You must select File | Save | Dark from the
menu to permanently save the spectrum to disk.

Note

If possible, do not turn off the light source when taking a dark spectrum. If you must turn
off your light source to store a dark spectrum, allow enough time for the lamp to warm up
again before continuing your experiment.

You must take a dark spectrum before measuring transmission.

5. Put the sample is in place and ensure that the light path is clear. Then, take a reflection

measurement by clicking on the Transmission Mode icon on the toolbar or selecting Spectrum |
Transmission Mode
from the menu bar (since the mathematical calculations used to calculate
transmission and reflection are identical). To save the spectrum to disk, click the Save icon on the
toolbar or select File | Save | Processed from the menu bar.

Note

If you change any sampling variable (integration time, averaging, smoothing, fiber size,
etc.), you must store a new dark and reference spectrum.