beautypg.com

18 $aa3nvv – Measurement Computing CB-COM-7021 User Manual

Page 49

background image

49

CB-7021, CB-7022, CB-7024 User’s Manual

2.18 $AA3NVV

Description:

Trim Calibration for Channel N

Note

: The command is for CB-7022 and CB-7024 only

Syntax:

$AA3NVV[CHK](cr)

$

Delimiter character

AA

Address of setting module (00 to FF)

3

Command for trimming calibration

N

Channel to trim (0 to 1 for CB-7022, 0 to 3 for

CB-7024)

VV

2’s complement hex to trim output value. 00 to 5F to

increase 0 to 95 counts, and FF to A1 to decrease 1 to 95
counts. A count = 4.88µA/2.44mV for CB-7022 and
2.44µA or 1.22mV for CB-7024.

Response:

Valid Command:

!AA[CHK](cr)

Invalid Command:

?AA[CHK](cr)

Syntax or comm. error may not get response.

!

Delimiter for valid command

?

Delimiter for invalid command

AA

Address of response module (00 to FF)

Example:
Command: $01321F

Receive: !01

Trim address 01 channel 2 output 31 counts, returns OK

Related Command

:

Sec. 2.16

$AA0N, Sec. 2.17 $AA1N

Related Topics

:

Sec. 1.9

Calibration

This manual is related to the following products: