17

CB-7021, CB-7022, CB-7024 User’s Manual

CB-7021/21P Voltage Output Calibration Sequence:
1.

Connect voltmeter to module’s voltage output.

2.

Warm-Up for 30 minutes.

3.

Set type to 32. (0 to 10V)

Refer Sec. 2.1.

4.

Program output for 10V.

Refer Sec. 2.7.

5.

Check

Refer Sec. 2.13.