CB-7024 Current Output Calibration Sequence:
1.

Connect meter to module’s voltage output channel 0.

2.

Warm-Up for 30 minutes.

3.

Setting type to 33. (

10 to 10V)

-> Refer Sec. 2.1.

4.

Output

10V.

-> Refer Sec. 2.15.

5.

Check the meter and trim the output for

10V match by using trim command.

-> Refer Sec. 2.18.

6.

Perform

10V Calibration Command. -> Refer Sec. 2.16.

7.

Program output for 10V.

Refer Sec. 2.15.

8.

Check the meter and trim the output for

10V match by using trim command.->

Refer

Sec. 2.18.

9.

Perform 10V Calibration Command.

-> Refer Sec. 2.17.

10. Repeat 1 to 9 for channel 1, 2 and 3.