CB-7024 Current Output Calibration Sequence:
1.

Connect meter and external power source to module’ s

current output channel 0.

2.

Warm-Up for 30 minutes.

3.

Set type to 30. (0 to 20 mA)

-> Refer Sec. 2.1.

4.

Program output for 0 mA.

-> Refer Sec. 2.15.

5.

Check the meter and trim the output for

0 mA match by using trim command.

-> Refer Sec. 2.18.

6.

Perform 0 mA Calibration Command. -> Refer Sec. 2.16.

7.

Program output for 20 mA.

-> Refer Sec. 2.15.

8.

Check the meter and trim the output for

20 mA match by using trim command. -> Refer Sec. 2.18.

9.

Perform 20 mA Calibration Command. -> Refer Sec. 2.17.

10. Repeat 1 to 9 for channel 1, 2, and 3.