Measurement Computing CB-COM-7021 User Manual
Page 20

20
CB-7021, CB-7022, CB-7024 User’s Manual
CB-7024 Current Output Calibration Sequence:
1.
Connect meter and external power source to module’ s
current output channel 0.
2.
Warm-Up for 30 minutes.
3.
Set type to 30. (0 to 20 mA)
-> Refer Sec. 2.1.
4.
Program output for 0 mA.
-> Refer Sec. 2.15.
5.
Check the meter and trim the output for
0 mA match by using trim command.
-> Refer Sec. 2.18.
6.
Perform 0 mA Calibration Command. -> Refer Sec. 2.16.
7.
Program output for 20 mA.
-> Refer Sec. 2.15.
8.
Check the meter and trim the output for
20 mA match by using trim command. -> Refer Sec. 2.18.
9.
Perform 20 mA Calibration Command. -> Refer Sec. 2.17.
10. Repeat 1 to 9 for channel 1, 2, and 3.