beautypg.com

17 $aa1n – Measurement Computing CB-COM-7021 User Manual

Page 48

background image

48

CB-7021, CB-7022, CB-7024 User’s Manual

2.17 $AA1N

Description:
CB-7022: Perform 20 mA Calibration for Channel N
CB-7024: Perform 20 mA/+10V Calibration for Channel N

Note

: The command is for CB-7022 and CB-7024 only.

Syntax:

$AA1N[CHK](cr)

$

Delimiter character

AA

Address of setting module (00 to FF)

1

Command for perform 20 mA (or +10V) calibration

N

Channel to calibrate (0 to 1 for CB-7022, 0 to 3 for

CB-7024)

Response:

Valid Command:

!AA[CHK](cr)

Invalid Command:

?AA[CHK](cr)

Syntax error or communication error may get no
response.

!

Delimiter for valid command

?

Delimiter for invalid command

AA

Address of response module (00 to FF)

Example:
Command: $0112

Receive: !01

Perform address 01 channel 1 calibration (20 mA for

CB-7022, 20 mA or 10.0V for CB-7024), returns successful.

Related Command

:

Sec. 2.16

$AA0N, Sec. 2.18 $AA3NVV

Related Topics

:

Sec. 1.9

Calibration

This manual is related to the following products: