AMETEK Ls Series User Manual
Page 80
User Manual
Lx / Ls Series
76
4.2.10 CALIBRATION Menus
The measurement calibration menu can be used to perform routine calibration of the internal
measurement system. The recommended calibration interval is 12 months. To enter the
calibration screens, the calibration password must be entered first.
Note: Refer to chapter 6 for details on routine calibration procedures and equipment
requirements. Do not attempt calibration without consulting the user manual.
This menu also contains the LCD viewing angle adjustment.
ENTRY
DESCRIPTION
VIEW ANGLE
-10 to +10
LCD viewing angle adjustment.
CAL PWORD
V range
Calibration password required to access all
calibration screens. The calibration password
is the high voltage range value. [300 ] The
password can be entered using the Voltage
shuttle followed by the ENTER key.
Measurement Calibration Screens
MVOLT F/S
± 9999
Calibration coefficient for full-scale voltage
measurement.
MCURR F/S
± 9999
Calibration coefficient for full-scale current
measurement.
Output Calibration Screens
VOLT F/S
± 9999
Calibration coefficient for full-scale voltage
output. (calibrated by measurement)
CURR F/S
± 9999
Calibration coefficient for full-scale current
ouptut. (calibrated by measurement)