2.1.3 AC Measurements

Measurement specifications apply to single chassis Lx / Ls Series AC source in single or three-
phase mode. See notes for other models and configurations.

Parameter

Range

Accuracy ( )

Resolution

1 Phase Mode

3 Phase Mode

Frequency

1

45.00-1000.0 Hz

0.1% ± 1 digit

0.01 Hz to 81.91 Hz
0.1 Hz to 819.1 Hz
1 Hz > 819.1 Hz

Phase

45.00 - 100.0 Hz
100.0 - 1000 Hz

0.5°

0. 1°

RMS Voltage

0 - 400 Volts

0.05% + 0.25V

0.01 Volt

RMS Current

0 - 50 Amps

0.1% + 0.15A

0.1% + 0.05A

0.001 Amp

Peak Current

0 - 150 Amps

0.2% + 0. 5A

0.1% + 0.15A

0.001 Amp

Crest Factor

1.00

– 10.00

1.5 %

1.5 %

0.01

VA Power

0 - 6 KVA

0.15% + 9 VA

0.15% + 3 VA

1 VA

Real Power

0 - 6 KW

0.15% + 9 W

0.15% + 3 W

1 W

Power Factor

0.00 - 1.00

0.03

0.01

0.01

Note: Accuracy specifications are valid above 100 counts. For multi-chassis configurations, Current
and Power range and accuracy specifications are times the number of chassis.

Note: Frequency measurement specification valid for output > 20 Vrms.

Note: Crest Factor accuracy applies for Irms > 50% of max.

Note: Power Factor accuracy applies for PF > 0.5 and VA > 50% of max.

2.1.4 Harmonic Measurements

Harmonic measurement specifications apply to Lx Series or Ls Series

–3 with –ADV option AC

sources in three-phase mode. See notes for single-phase mode or Ls

–1 with –ADV option.

Parameter

Range

Accuracy ( )

Resolution

Frequency fundamental

45.00 - 81.91 Hz
82.0 - 819.1 Hz
> 819.1 Hz

0.1% ± 1 digit

0.01 Hz
0.1 Hz
1 Hz

Frequency harmonics

45.00 Hz

– 16 kHz

0.1% + 2 digits

0.1 Hz

Voltage

Fundamental 0 - 400 Volts

0.05% + 0.25V

0.01V

Harmonic 2 - 50

0.1% + 0.1%/kHz + 0.25

0.01V

Current

Fundamental 0 - 20 Amps

0.1% + 0.05A

0.01A

Harmonic 2 - 50

0.1% + 0.1%/kHz + 0.05A

0.01A

Note: Current range and accuracy specifications are times three in single-phase mode. For multi-
chassis configurations, current accuracy specifications are times the number of chassis.

1

Frequency measurement specifications valid with output voltage of 30Vrms or higher. If output relay is open,

frequency measurement will return 0.0 Hz.