Table 3-4 – AMETEK Ls Series User Manual
Page 50

User Manual
Lx / Ls Series
46
SMA
Description
OUT1
(Can be set to Function Strobe for compatibility with CI L Series)
TRIGGER
IN1
Trigger Input (TTL input)
RPV
Not Used.
Table 3-4: SMA Connectors
– Lx Series
3.6.3 SMA Connectors
– Ls Series
SMA connectors. Functions are called out on rear panel decal. Table 3-5 shows connections
from left to right when standing at the rear of the Ls cabinet. SMA connectors are small high
frequency capable coax connectors that can be screwed down securely to prevent signal loss.
Adaptor for SMA to BNC conversion are readily available.
To connect these signals using more conventional BNC cables, SMA Male to BNC Female
adapters may be used. These are available from
PE9074 Adaptor.
SMA
Description
J3
DFI
Discrete Fault Indicator output. Isolated Open Collector. Can be used to signal
external devices when a fault condition is detected.
J7
36/27
INH
Remote Shutdown. (Inhibit) (TTL input) Equivalent to J7-36 (+) and J7-27 (-) on CI L
Series.
J7
31/14
FSTB
Function Strobe
(Can be set to Trigger Output (TTL output) for compatibility with CI iL Series)
Equivalent to J7-31 (+) and J7-14 (-) on CI L Series.
TRIG
Trigger Input (TTL input)
RPV
Not Used.
Table 3-5: SMA Connectors
– Ls Series