beautypg.com

3 measurement calibration – AMETEK Ls Series User Manual

Page 100

background image

User Manual

Lx / Ls Series

96

6.3 Measurement Calibration

The Lx/Ls Series controller measures voltage and current by digitizing both voltage and current
waveforms on each available output phase. This data is subsequently processed and use to
calculate all measurement parameters such as VRMS, IRMS, Power, VA, Frequency etc. To
calibrate all measurements, only the voltage and current measurement need to be calibrated
specifically. All other measurements are derived from these.

Connect the test equipment to the power source. If the power system is a master/auxiliary multi-
box system with one controller, the DVM for calibrating the measurement voltage should always
be connected to the Remote Sense connector on the Master cabinet.

Note: The Fluke 8506A or Agilent HP 34401A Digital Multi meter (or higher AC accuracy

DMM) must be used for the following calibration. The DMM must be set to the AC
HI ACCUR mode for all AC measurements.

The shunt must be connected in series with the load. Connect the load to the output. Use a 10
mOhm current shunt of sufficient power rating in series with the load to measure the AC load
current.

To calibrate all measurement functions, the desired value for the measurement value of current
or voltage must be entered for the corresponding calibration value. Make the indicated
adjustments by typing in the desired display value. This should be the value indicated by the
external DVM. If a 10 mOhm current shunt is used for current, 300 mV represents 30 amps.

The Calibration Load Table shows required load bank settings for the current measurement
calibration procedure. The current should be calibrated in the lowest voltage range only. (Highest
current range). The current measurement calibration must be performed for Phase A in both
single and three phase mode and for phase B and C in three phase mode only.

PARAMETER

POWER SYSTEM

Model --->

3000Lx

1 Phs

3000Lx

3 Phs

4500Lx

1 Phs

4500Lx

3 Phs

6000Lx

1 Phs

6000Lx

3 Phs

Max current, 120 V, Lo
Vrange

4.8

3 kW

14.4

1 kW

3.2

4.5 kW

9.6

1.5 kW

2.4

6 kW

7.2

6 kW

Table 6-1: Calibration Load Values- Single-chassis configurations

PARAMETER

POWER SYSTEM

Model --->

9000Lx/2

1 phs mode

9000Lx/2

3 phs mode

12000Lx/2

1 phs mode

12000Lx/2

3 phs mode

Max current, 120 V, Lo
Vrange

1.6

9 kW

4.8

3 kW

1.2

12 kW

3.6

4 kW

Model --->

13500Lx/3

1 phs mode

13500Lx/3

3 phs mode

18000Lx/2

1 phs mode

18000Lx/3

3 phs mode

Max current, 120 V, Lo
Vrange

0.96

13.5 kW

3.2

4.5 kW

0.8

18 kW

7.2

6 kW

Table 6-2: Calibration Load Values- Multi-chassis configurations

This manual is related to the following products: