AMETEK Ls Series User Manual
Page 188
User Manual
Lx / Ls Series
184
SXF 401 Tests
– Revision F
SXF401
Emergency SS Limits Volt and Freq. Rev F
Condition
Duration (s)
Nominal Voltage
A
Vnom=115V, Fnom=60 Hz - Nominal
1800
B
Vnom=115V, Flow=59.5 Hz - Nominal
1800
C
Vnom=115V, Fhigh=60.5 Hz - Nominal
Low Voltage
D
Vlow =105V, Fnom=60 Hz - Low
1800
E
Vlow=105V, Flow= 59.5 Hz- Low
1800
F
Vlow=105V, Fhigh=60.5 Hz - Low
1800
High Voltage
G
Vhigh=125V, Fnom=60 Hz - High
1800
H
Vhigh=125V, Flow=59.5 Hz- High
1800
I
Vhigh=125V, Fhigh=60.5 Hz - High
1800
SXF 501 Tests
Starting operations are not applicable to AC Utilization Equipment.
SXF 601 Tests
– Revision F
SXF601
Power Failure - One Phase
Condition
Duration (s)
A
100 msec
52
B
500 msec
54
C
2 sec
66
SXF 602 Tests
– Revision F
Not applicable for single-
phase EUT’s.
SXF 603 Tests
– Revision F
SXF603
Power Failure
– Phase Reversal
Condition
Duration (s)
N/A
Use physical L-N connection reversal
Apply Vnom = 115 V, Fnom = 60 Hz
1800