3 confirming an active alarm – BINDER KT 53 User Manual
Page 93

KT (E6) 09/2013
Page 93/135
14.3 Confirming an active alarm
Fixed value
08.03.2013 15:05:02
Initial view with overtemperature safety controller
alarm.
The buzzer sounds (if not deactivated previously).
Press the “Info” button.
Setpoint
Actual value
Temperature [°C]
25.0 29.8
Fan
[%]
100.0
User
Info
Menu
Alarm acknowledgement
08.03.2013 05:05:06
Submenu “Alarms” for alarm acknowledgement.
The active alarm that has been triggered first is
highlighted.
Press the “Reset” button to confirm the alarm and
mute the buzzer.
..\ Home/Alarms
Safety controller overtemperature
Close
Reset
Home
The initial view is displayed. The alarm is still active, therefore the alarm icon continues to be shown. The
buzzer is off.
If there is another active alarm, you can press the “Info” button again and also confirm the alarm with the
“Reset” button.
The following alarm messages may appear in the “Alarm list”:
Alarm message
Signification
Corresponding
alarm icon
Safety controller
overtemperature
Overtemperature safety controller alarm (class 3.1): selected
value of the safety controller exceeded.
Safety controller
overtemperature
Overtemperature safety controller alarm with option class 3.3:
selected value of the safety controller exceeded
Safety controller
undertemperature
Undertemperature safety controller alarm with option class 3.3:
fallen below selected value of the safety controller
Temp. range
Having reached the set-point, the temperature deviates by
more than +/- 2 °C from the entered set-point