8.
Set-point entry in “Fixed value” operating mode
8.1 Setting ranges:
Temperature
4 °C / 23 °F up to +100 °C / 212 °F
Fan speed
40 % up to 100 % (full speed)
Reduce the fan speed only if required, because the spatial distribution of temperature
will also be reduced.
Technical data refers to 100% fan speed.
When you changed the temperature set-point, check the setting of the overtemperature safety
controller class 3.1 (chap. 17.2) or the over-/undertemperature safety controller class 3.3
(option, chap. 17.3).
With set-point type “Limit”, adapt the safety controller always when changing the temperature
set-point.
8.2 Entering the set-points via “quick menu”
To enter set-points via quick menu, go to
Quick menu
.
Fixed value
08.03.2013 05:05:06
“Quick menu”.
Select the desired parameter
and press the operating button.
..\ Quick menu
Measurement chart
Active alarms
Temperature setpoint
Fan speed setpoint
Safety controller setpoint
Time program
Week program
Close
Home
When trying to enter a setpoint in the „Controller off“ operating mode, a notification window
shows “Controller mode is OFF!”. Press the operating button to confirm with “Ok” and change
the operating mode to “Fixed value” (chap. 6.2.1).