beautypg.com

Configuration of optional equipment, 1 setting the optional door heating, 5 [°c – BINDER KT 53 User Manual

Page 31

background image

KT (E6) 09/2013

Page 31/135

7.

Configuration of optional equipment

The “Optional equipment” menu item is visible only with optional unit equipment.

To access the selection menu, go to

Menu > Optional equipment

Fixed value

08.03.2013 05:05:06

Submenu “Optional equipment”.
Select the desired function

and press the operating button.

..\ Optional equipment

Door heating

Interior socket

Zero-voltage relay control outputs

Alarm output temperature

Object temperature display

Close

Home

7.1 Setting the optional door heating

For chambers equipped with an optional door heating, you can turn it on and off via the controller. You

can also set the door heating offset to the temperature set-point.
To access the door heating setting menu, go to

Menu > Optional equipment > Door heating

Fixed value

08.03.2013 05:05:06

Submenu “Door heating”.
Select the desired function

and press the operating button.

Door heating On = door heating turned on

Door heating Off = door heating turned off

Door heating Offset = Temperature difference to the

entered setpoint

..\ Door heating On/Off

Door heating On

Door heating Off

Door heating Offset

Close

Home


Setting the offset value:

Fixed value

08.03.2013 05:05:06

Entry menu „Offset“.
Select each number with the operating button and

press the operating button to confirm.
Setting range: 0 °C up to 5 °C.
Press the “Ok” button to confirm.

..\ Door heating \ Offset

1

.5

[°C]

0

Ins Pos1 End Ok

1 2 3 4 5 6 7 8 9 ,

Close

Ok

Home