BINDER KT 53 User Manual
Page 55
KT (E6) 09/2013
Page 55/135
Entering the set-point values for another parameter (fan speed)
The number of program lines (program sections) equal to the number in the temperature program is
displayed. The settings of section length, repeats and the selection „Ramp“ or „Step“ are taken over from
the temperature program; they are no editable in this view. The symbol “ *** ” indicates that there are no
tolerance value required for the fan speed. You can enter the fan speed set-points.
If you want to insert further sections or perform any other programming, first change back to
the temperature program through
Menu > Select parameter > Temperature controller
.
Fixed value
08.03.2013 05:05:06
Time program editor (example).
To edit the values, press the operating button, select
the desired value and press the operating button
again.
..\ Fan speed
[%]
No. Value H:M:S Ref. Rep. T. min T.max R/S
1 100.00 00:30:00 1
0
***
*** Ramp
1 100.00 00:10:00 1
0
***
***
Ramp
1 100.00 00:30:00 2
3
***
***
Ramp
1 100.00 00:00:01 1
0
***
***
Ramp
Menu
Entering the fan speed setpoint
Fixed value
08.03.2013 05:05:06
Time program editor.
Select a value under “Value”
and press the operating button.
..\ Fan speed
[%]
No. Value H:M:S Ref. Rep. T. min T.max R/S
1
100.00
00:00:15 1
0
***
***
Ramp
Fixed value
08.03.2013 05:05:06
Entry menu “Fan speed setpoint”.
A fan speed value is shown. Enter the desired value
with the operating button.
Setting range: 40 % up to 100 %
Press the “Ok” button to confirm.
..\ Fan speed setpoint
1
00
[%]
1
Pos1 End Ok 0
2 3 4 5 6 7 8 9 ,
Close
Ok
When entering a value outside the setting range, the message “invalid value” appears. Press the
operating button to confirm with “Ok” and repeat the entry with a correct value.