beautypg.com

2 temperature setpoint, 3 section duration, 00 [°c – BINDER KT 53 User Manual

Page 47

background image

KT (E6) 09/2013

Page 47/135

9.3.2 Temperature setpoint

Fixed value

08.03.2013 05:05:06

Time program editor.
Select a value under “Value”

and press the operating button.

..\ Temperature controller

[°C]

No. Value H:M:S Ref. Rep. T. min T.max R/S

1

25.00

00:00:15 1

0 -999.00 999.00 Ramp

08.03.2013 05:05:06

Entry menu “Temperature setpoint”.
A temperature value is shown. Enter the desired

value with the operating button.
Setting range: 4 °C up to 100 °C.
Press the “Ok” button to confirm.

..\ Temperature setpoint

2

5.00

[°C]

2

Pos1 End Ok 0 1

3 4 5 6 7 8 9 , -

Close

Ok

When entering a value outside the setting range, the message “invalid value” appears. Press the

operating button to confirm with “Ok” and repeat the entry with a correct value.

9.3.3 Section duration

Fixed value

08.03.2013 05:05:06

Time program editor.
Select a value under “H:M:S”

and press the operating button.

..\ Temperature controller

[°C]

No. Value H:M:S Ref. Rep. T. min T.max R/S

1 25.00

00:00:15

1

0 -999.00 999.00 Ramp

Fixed value

08.03.2013 05:05:06

Entry menu “Section duration”.
A duration value is shown. Enter the desired value

with the operating button.
Press the “Ok” button to confirm.

..\ Section duration

0

0:00:15

0

Ins Pos1 End Ok

1 2 3 4 5 6 7 8 9

Close

Ok

This manual is related to the following products: