4 kt technical data – BINDER KT 53 User Manual
Page 125
KT (E6) 09/2013
Page 125/135
23.4 KT technical data
Unit size
53
115
Exterior dimensions
Width
mm / inch
660 / 25.98
860 / 33.86
Height (incl. feet)
mm / inch
635 /
715 / 28.15
Depth (incl. door handle, instrument triangle,
connection)
mm / inch
695 /
720 / 28.35
Wall clearance rear (minimum) (spacer)
mm / inch
100 / 3.94
100 / 3.94
Wall clearance side (minimum)
mm / inch
240 / 9.45
240 / 9.45
Quantity of doors
1
1
Quantity of inner glass doors
1
1
Interior dimensions
Width
mm / inch
400 / 15.75
600 / 23.62
Height
mm / inch
400 / 15.75
480 / 18.90
Depth
mm / inch
334 / 13.15
355 / 14.00
Interior volume
l / cu.ft.
52 / 1.84
102 / 3.60
Quantity of racks standard / max.
2 / 5
2 / 6
Max. load per rack
Kg / lbs
15 / 33.1
40 / 88.2
Permitted total load
Kg / lbs
40 / 88.2
120 / 264.5
Weight (empty)
Kg / lbs
65 / 143.3
88 / 194.0
Temperature data
Temperature range 1)
°C / °F
4 °C / 23 °F up to +100 °C / 212 °F
Temperature fluctuation
max.
± K
0.1
0.1
Temperature uniformity
(variation)
max.
± K
2.0
2.0
at 4 °C / 39.2°F
± K
0.3
0.3
at 25 °C / 77°F
± K
0.1
0.1
at 37 °C / 98.6°F
± K
0.3
0.3
Max. heat compensation up to 40 °C / 104°F
W
100
100
Recovery time after doors
were open for 30 sec. 2)
at 4 °C / 39.2°F
minutes
8
8
at 25 °C / 77°F
minutes
1
1
at 37 °C / 98.6°F
minutes
1
1
Electrical data
IP System of protection acc. to EN 60529
20
20
Nominal voltage (+/-10%)
V
200-240 1N~
200-240 1N~
Power frequency
Hz
50/60
50/60
Power plug (IEC connector plug)
Shock proof plug
Nominal power
kW
1.10
0.70
Energy consumption 3)
at 4 °C / 39.2°F
Wh/h
250
250
at 25 °C / 77°F
Wh/h
90
90
at 37 °C / 98.6°F
Wh/h
75
75
Overvoltage category acc. to IEC 61010-1
II
II
Pollution degree acc. to IEC 61010-1
2
2
Unit fuse 5x20 mm 230V / middle-time-lag M
Amp
10 external
10 external
Noise level (approx.)
dB (A)
***
***