beautypg.com

BINDER KT 53 User Manual

Page 56

background image

KT (E6) 09/2013

Page 56/135

9.3.9 Saving the time program and leaving the program editor

Fixed value

08.03.2013 05:05:06

Time program editor (example).
Press the “Menu” button.

..\ Temperature controller

[°C]

No. Value H:M:S Ref. Rep. T. min T.max R/S

1 40.00 00:30:00 1

0 -999.00 999.00 Step

2 60.00 01:30:00 1

0

-5.00 5.00 Step

3 80.00 01:00:00 1

0

-5.00 5.00 Step

4 20.00 03:20:00 1

0 -999.00 999.00 Step

5 40.00 00:30:00 1

0 -999.00 999.00 Step

6 60.00 01:30:00 1

0

-5.00 5.00 Step

7 70.00 01:00:00 1

0

-5.00 5.00 Step

8 60.00 01:00:00 1

0

-5.00 5.00 Step

Menu

Fixed value

08.03.2013 05:05:06

Submenu “Program menu”.
Select “Save and exit”

and press the operating button.

..\ Program menu
Select parameter

Save and exit

Save and run

Exit (without saving!)

Close

With the „Close“ button, the controller returns to the time program editor.

Select parameter

Changing between temperature and fan speed.

Save and exit

Saving the program. The controller returns to the “Time programs”

submenu. You can now select and start the program as described

in chap. 9.1.

Save and run

Saving the program and start it. Enter the start date and time, see

chap. 9.1. If another time or week program is running, the program

is only saved, but not started. A corresponding message is

displayed.

Exit (without saving!)

Attention: the program is not saved. After a security question, the

controller returns to the initial view.

Button “Close”

The controller returns to the time program editor. You can continue

programming.

Make sure that you saved the time program before leaving the program editor.

This manual is related to the following products: