2 cleaning and decontamination, 1 cleaning, Danger – BINDER KT 53 User Manual
Page 116: Caution

KT (E6) 09/2013
Page 116/135
20.2 Cleaning and decontamination
Clean the unit after each use to avoid potential corrosion damage by ingredients of the test material.
DANGER
Electrical hazard.
Danger of death.
∅
Do NOT spill water or cleaning agents over the inner and outer surfaces.
Before cleaning, turn off the unit at the main power switch and disconnect
the power plug.
Completely dry the appliance before turning it on again.
20.2.1 Cleaning
Disconnect the chamber from the power supply before cleaning. Disconnect the power plug.
Wipe the surfaces with a moistened towel. In addition, you can use the following cleaning agents:
Exterior surfaces
inner chamber
shelves
door gaskets
Standard commercial cleaning detergents free from acid or halides.
Alcohol based solutions.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
Instrument panel
Standard commercial cleaning detergents free from acid or halides.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
Zinc coated hinge parts
rear unit wall
Standard commercial cleaning detergents free from acid or halides.
Do NOT use a neutral cleaning agent on zinc coated surfaces.
We recommend using the neutral cleaning agent Art. No. Art. Nr. 1002-0016 for thorough
cleaning.
Any corrosive damage that may arise following use of other cleaning agents is excluded from
liability by BINDER GmbH.
Any corrosive damage caused by a lack of cleaning, is excluded from liability by BINDER
GmbH.
CAUTION
Danger of corrosion.
Damage to the unit.
∅
Do NOT use acidic or chlorine cleaning detergents.
∅
Do NOT use a neutral cleaning agent on other kind of surfaces e.g., the zinc coated
hinge parts or the rear unit wall.
For surface protection, perform cleaning as quickly as possible.
After cleaning completely remove any cleaning agents from the surfaces by using a moistened
towel. Let the unit dry.
Soapsuds may contain chlorides and must therefore NOT be used for cleaning.