12.5 Changing the temperature unit

To select the temperature unit, go to

Menu > Settings > Temperature unit

Fixed value

08.03.2013 05:05:06

Submenu “Temperature unit”.
Select the desired temperature unit”

and press the operating button.

..\ Temperature unit

Celsius [°C]

Fahrenheit [°F]

Kelvin [K]

Close

Home

Go back to the “Settings” menu with “Close” or to the initial view with “Home”.

12.6 Defining the data recording rate

To define the recording rate for data storage on the SD card, go to

Menu > Settings > Recording rate

Fixed value

08.03.2013 05:05:06

Entry menu “Recording rate”.
The current recording rate is shown. Enter the

desired value with the operating button.
Shortest possible value: 60 seconds.
Press the “Ok” button to confirm.

..\ Recording rate (seconds)

6

0

0

Ok 0 1 2 3 4 5

7 8 9 . – Del Ins

Close

Ok

Home

Go back to the “Settings” menu with “Close” or to the initial view with “Home”.

Note:
After a period of 9 years, the controller starts overwriting the oldest values on the storage medium. This is

independent of the selected storage interval and the actual operating time of the chamber. In any case,

the data can be read out at any time using the function “Export to USB drive” (chap. 13.1) and stored

externally.