beautypg.com

0 [°c – BINDER KT 53 User Manual

Page 35

background image

KT (E6) 09/2013

Page 35/135


Temperature setting
To enter the temperature setpoint, go to

Quick menu > Temperature

setpoint

Fixed value

08.03.2013 05:05:06

Entry menu “Temperature setpoint”.
Select each number with the operating button and

press the operating button to confirm.
Setting range: 4 °C / 23 °F up to +100 °C / 212 °F
Press the “Ok” button to confirm.

..\ Temperature setpoint

2

5.0

[°C]

2

End Ok 0 1

3 4 5 6 7 8 9 . -

Close

Ok

Home

When entering a value outside the setting range, the message “invalid value” appears. Press the

operating button to confirm with “Ok” and repeat the entry with a correct value.
Go back to the initial view with “Home” or enter the fan speed.

Fan speed setting
To enter the fan speed setpoint, go to

Quick menu > Fan speed setpoint

Fixed value

08.03.2013 05:05:06

Entry menu “Fan speed setpoint”.
Select each number with the operating button and

press the operating button to confirm.
Setting range: 40 % up to 100 %
Press the “Ok” button to confirm.

..\ Fan speed setpoint

1

00

[%]

1

Pos1 End Ok 0

2 3 4 5 6 7 8 9 , -

Close

Ok

Home

When entering a value outside the setting range, the message “invalid value” appears. Press the

operating button to confirm with “Ok” and repeat the entry with a correct value.
Go back to the initial view with “Home”.

When operating the fan with less than 100 % speed, the temperature performance and the

spatial exactitude of the temperature may differ from the manufacturer’s specifications. Do

reduce the fan speed only if absolutely necessary due to special requirements.

This manual is related to the following products: