19 info - and error messages – ALTANA Wave-Scan II User Manual
Page 57

56
English
19
Info - and Error Messages
Identifies an internal error. Load a
firmware update. When in doubt, call
the Customer Service department.
The measurement unit was moved
across the sample too fast. The
measurement is invalid and must be
repeated.
The instrument was accelerated too
fast across the specimen or the scan
direction was changed too fast.
Confirm the message by pressing
“operate” and repeat the
measurement.
Appears when the „Plausibility
Control“ function in the Configuration
of the used Memory or Organizer file is
ON and the corrected value deviates
by more than 20% from the
uncorrected value. Repeat the
measurement.
Not enough light is reflected by the
surface. Repeat the measurement on
an appropriate sample area.
The operate key was released before
the full scanlength was reached. Per-
form a new measurement.
Day or month falls outside the valid
range of 1 - 31 or 1 - 12. Repeat the
entry.
Error Messages