Configuration 10 – ALTANA Wave-Scan II User Manual
Page 32

31
English
Configuration
10
Number of Measurements
This function allows selection of the
numbers of readings to be taken per
sample. If the number is greater than
1, the measurements are statistically
evaluated. Turning the scroll wheel
adjusts the number while pressing it
shifts the decimal place one further
over. Pressing operate completes the
process and a confirmation display
appears.
Scanlength
The setting options for Scanlength are
shown in the neighboring illustration.
The current setting is identified by a
black dot at the end of the line.
Move the cursor to the desired setting
and press the operate button. The
instrument returns to the previous
menu.
Note:
Using a short scanlength will result in
strong statistical skewing of
measurement values for large
wavelengths. Thus, when using the We
or Wd scale with a scanlength of
5 cm, at least 3 measurements should
be made per checkzone to ensure
representative values.